Equipment Entry

List of Maintenance Log Entries

Next PM Date

{some date}

Has Notified Customer

{true / false}

Scheduled PM Dates

{re-schedule date #2}, {re-schedule date #1}, {schedule date}

Actual PM Date

2099-09-09

Notes

9Jan2022: Wai Chung informed that they want PM of the GCMS (Model: 6890N, SN: 5973) and Frontier Multi-shot Pyrolyzer (Model: PY-3030D, SN: PY1522320291D) ONCE IN EVERY 2 YRS i/o annually. NOTE: Last PM was on 18 Jan 2021 & therefore will skip Yr 2022 PM to next year due 18 Jan 2023. To follow up with him in Dec 2022.

Last modified

2026-03-18T09:11:47.383512Z

Next PM Date

{some date}

Has Notified Customer

{true / false}

Scheduled PM Dates

{re-schedule date #2}, {re-schedule date #1}, {schedule date}

Actual PM Date

2024-02-21

Notes

J04775: Repair Frontier Multi-shot Pyrolyzer. Dismantling & installation completed with styrene run ok.

Last modified

2026-03-18T09:11:47.383482Z

Next PM Date

{some date}

Has Notified Customer

{true / false}

Scheduled PM Dates

{re-schedule date #2}, {re-schedule date #1}, {schedule date}

Actual PM Date

2023-11-17

Notes

J04767 (=> PO#4700218587) : Replaced all PM parts provided by user. Leaking issue resolved.

Last modified

2026-03-18T09:11:47.383454Z

Next PM Date

{some date}

Has Notified Customer

{true / false}

Scheduled PM Dates

{re-schedule date #2}, {re-schedule date #1}, {schedule date}

Actual PM Date

2023-01-26

Notes

J04688 (=>4700184622) : PM completed as per PM checklist

Last modified

2026-03-18T09:11:47.383424Z

Next PM Date

{some date}

Has Notified Customer

{true / false}

Scheduled PM Dates

{re-schedule date #2}, {re-schedule date #1}, {schedule date}

Actual PM Date

2023-01-17

Notes

PO rcvd : 4700184622 (=>QT-202301009) : PM for Frontier Multi-shot Pyrolyzer (SN: PY1522320291D) & GCMS (SN: 5973)

Last modified

2026-03-18T09:11:47.383376Z

Next PM Date

{some date}

Has Notified Customer

{true / false}

Scheduled PM Dates

{re-schedule date #2}, {re-schedule date #1}, {schedule date}

Actual PM Date

2023-01-06

Notes

QT-202301009 : PM for Frontier Multi-shot Pyrolyzer (SN: PY1522320291D) & GCMS (SN: 5973)

Last modified

2026-03-18T09:11:47.383348Z

Next PM Date

{some date}

Has Notified Customer

{true / false}

Scheduled PM Dates

{re-schedule date #2}, {re-schedule date #1}, {schedule date}

Actual PM Date

2021-01-18

Notes

J04249: Pyrolyzer with GCMS PM completed. Checking AED System condition. GCMS PM Kit replaced & Ion source cleaning completed. Pyrolyzer (PY15223202910) PM completed as per PM checklist. AED unable to communicate with software & plasma unable to power up. There is a problem at either power supply to the mainboard or the magnetron is faulty. Check main board test point as per manual unable to achieve required reading.

Last modified

2026-03-18T09:11:47.383320Z

Next PM Date

{some date}

Has Notified Customer

{true / false}

Scheduled PM Dates

{re-schedule date #2}, {re-schedule date #1}, {schedule date}

Actual PM Date

2021-01-07

Notes

PO rcvd: 4700096265 - PM for Frontier Multi-shot Pyrolyzer (SN: PY1522320291D) & GCMS (SN: 5973)

Last modified

2026-03-18T09:11:47.383292Z

Next PM Date

{some date}

Has Notified Customer

{true / false}

Scheduled PM Dates

{re-schedule date #2}, {re-schedule date #1}, {schedule date}

Actual PM Date

2021-01-04

Notes

QT-202007001R3: Annual PM for Frontier Multi-shot Pyrolyzer (PY1522320291D) and GCMS (6890-5973)

Last modified

2026-03-18T09:11:47.383263Z

Next PM Date

{some date}

Has Notified Customer

{true / false}

Scheduled PM Dates

{re-schedule date #2}, {re-schedule date #1}, {schedule date}

Actual PM Date

2020-09-17

Notes

J04112: Check Pyrolyzer furnance and ITF connection ok. Carry out basic maintenance for pyrolyzer. Replace ac liner and inlet septum. Tighten pyrolyzer internal connection and replace pyrolyzer inlet seal. Calibration of pyrolzer furnance temperature.

Last modified

2026-03-18T09:11:47.383235Z

Next PM Date

{some date}

Has Notified Customer

{true / false}

Scheduled PM Dates

{re-schedule date #2}, {re-schedule date #1}, {schedule date}

Actual PM Date

2020-09-11

Notes

PO rcvd: 4700082835 - Troubleshooting of Frontier Pyrolyzer - 2 hrs

Last modified

2026-03-18T09:11:47.383207Z

Next PM Date

{some date}

Has Notified Customer

{true / false}

Scheduled PM Dates

{re-schedule date #2}, {re-schedule date #1}, {schedule date}

Actual PM Date

2020-09-09

Notes

QT-202009021: Troubleshooting of Frontier Pyrolyzer - 2 hrs

Last modified

2026-03-18T09:11:47.383179Z

Next PM Date

{some date}

Has Notified Customer

{true / false}

Scheduled PM Dates

{re-schedule date #2}, {re-schedule date #1}, {schedule date}

Actual PM Date

2020-07-14

Notes

QT-202007001R1: PM for Frontier Multi-shot Pyrolyzer (Model: PY-3030D, SN: PY1522320291D) & GCMS (GCMS 6890-5973N)

Last modified

2026-03-18T09:11:47.383150Z

Next PM Date

{some date}

Has Notified Customer

{true / false}

Scheduled PM Dates

{re-schedule date #2}, {re-schedule date #1}, {schedule date}

Actual PM Date

2020-07-01

Notes

QT-202007001 : Annual PM for Frontier Multi-shot Pyrolyzer

Last modified